Study on the Detection Method of Hemodynamic Parametersin Different Stenosis
CSTR:
Author:
Affiliation:

Clc Number:

Q81

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    To investigate the hemodynamics parameter of stress phase angle (SPA) in different stenosis states, the high-frequency ultrasonic imaging system was employed to obtain ultrasound contrast images of stenotic plaques, and the value of SPA was calculated through echo particle image velocimetry and cross-correlation algorithms. The experimental results showed that when the stenosis was 30%, 50% and 70%, the mean SPA was -162.88°, -201.99° and -222.89°, respectively. The study may suggest that the bigger stenosis, the more negative the SPA and the more asynchronous between the wall shear stress and circumferential stress.

    Reference
    Related
    Cited by
Get Citation

LI Shuai, NIU Lili, QIAN Ming, et al. Study on the Detection Method of Hemodynamic Parametersin Different Stenosis[J]. Journal of Integration Technology,2016,5(4):81-87